| Model Number |
115-203 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture (grade iii).The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture (baker grade iii).The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: creases flat, wear abrasion, deformation and the weight the device within specification.Cloudy and voids in the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.
|
| |
|