C.R. BARD, INC. (COVINGTON) -1018233 BARD® NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER; NG TUBE
|
Back to Search Results |
|
Model Number 0046160 |
Device Problems
Difficult to Insert (1316); Deformation Due to Compressive Stress (2889); Material Deformation (2976)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
|
Event Description
|
It was reported that the ng tube was kinked at the end.The hospital had numerous difficulties with numerous patients.In addition, they had difficulty inserting them.The doctor was unable to place one of the ng tubes.Another doctor was able to place one after icing the first 6 cm or so, with some difficulty.The doctor stated that these tubes appear to have multiple openings, with one large one, and they didn't remember there being so many.
|
|
Manufacturer Narrative
|
The device was not returned for evaluation.The reported event could not be confirmed.A potential failure mode could be ¿tube kinks¿ with a potential root cause of ¿inadequate material selection during design phase¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿caution: read all instructions prior to use.Indications for use: bard nasogastric sump tubes are intended to be used for: ¿ decompression of stomach by suction or aspiration of gastric contents.¿ short-term administration of term tube feeding, lavage fl uid and medications.Contraindications: ¿ patients with known tape or adhesive allergies.Warnings 1.Use with caution in patients with a history of head trauma, facial trauma, esophageal diseases and patients with potential for vomiting.2.Do not force nasogastric tube during insertions; damage to the nasal passage and mucosa and bleeding may occur.3.Measure insertion length carefully- excessive insertion length of tube into the stomach may lead to coiling and/or formation of tube-knot.4.Lubricate the tube generously with water soluble lubricant prior to insertion.Do not use petroleum-based products as they may be harmful to the respiratory tract.5.Refl ux of gastric contents into the blue vent lumen indicates that the suction lumen is obstructed or suction is too low.Routinely check for refl ux in the blue vent lumen and clear as per applicable directions.Failure to clear the obstruction or clear prevent® filter may cause gas and fluid buildup in stomach, aspiration of gastric contents, aspiration pneumonia and other complications.6.Do not inject fl uid through the prevent® filter as this may result in blockage and leakage of fi lter.7.Monitor patient for nasal erosion, sinusitis, esophagitis, esophagotracheal fi stula, gastric erosion and pulmonary & oral infections." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was not returned.
|
|
Event Description
|
It was reported that the ng tube was kinked at the end.The hospital had numerous difficulties with numerous patients.In addition, they had difficulty inserting them.The doctor was unable to place one of the ng tubes.Another doctor was able to place one after icing the first 6 cm or so, with some difficulty.The doctor stated that these tubes appear to have multiple openings, with one large one, and they didn't remember there being so many.
|
|