Advanced bionics considers the investigation into this reportable event as closed.The recipient reportedly recovered.No further information will be provided.The external visual inspection revealed silicone slices on the top cover and the electrode was severed along the electrode and near the array prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|