U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751LNAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751LNAB; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751LNAB
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Device Difficult to Program or Calibrate (1496); Obstruction of Flow (2423); Material Integrity Problem (2978)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Information received by medtronic indicated that buttons were hard to press and not responsive.Customer reported that insulin pump had no delivery alarm,battery cap was worn and back light was hard.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device received with all buttons responding properly.No damage on keypad assembly.No unlocked lcd keypad connector.Device passed the displacement test, rewind, self test, basic occlusion, occlusion, prime/a33 and excessive no delivery test.No excessive no delivery or occlusion alarm noted.No unexpected back light anomalies noted.Device received with stripped battery cap coin slot.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751LNAB
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10006512
Report Number2032227-2020-111052
Device Sequence Number3406603
Product Code OZO
UDI-Device Identifier00643169933187
UDI-Public(01)00643169933187
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751LNAB
Device Catalogue NumberMMT-751LNAB
Device Lot NumberA5751LNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2020
Initial Date Received by Manufacturer 04/23/2020
Supplement Date Received by Manufacturer06/26/2020
Initial Report FDA Received Date04/28/2020
Supplement Report FDA Received Date06/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
Patient Weight229
-
-