| Model Number |
77000FD |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the enteral feeding set is leaking at the clear tubing.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation.An examination of the photograph detected a cross spike detached from the tubing line.The root cause and action plan will be documented through a formal corrective/preventative action which has been initiated to address the reported condition to mitigate any further occurrences.This action is currently ongoing.This complaint will be closed with no further action and will be used for tracking and trending purposes.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation.One used device was received for evaluation.A visual and functional inspection confirmed that the cross spike was detached from the tubing line.The root cause and action plan will be documented through a formal corrective/preventative action which has been initiated to address the reported condition to mitigate any further occurrences.This complaint will be closed with no further action.This complaint will be used for tracking and trending purposes.
|
| |
|