| Model Number |
MMT-723NAP |
| Medical Device Problem Codes |
No Display/Image (1183); Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had button error alarm.The customer stated insulin pump had issues with the buttons for about 1 year but this morning it was not working.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device monitored for several days and no blank display noted.Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device received with normal operating current.Device passed self test.Device passed off no power test.
|
| |
|
Search Alerts/Recalls
|