Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 8100673, medical device expiration date: 2023-05-31, device manufacture date: 2018-04-10.Medical device lot #: 8100674, medical device expiration date: 2023-05-31, device manufacture date: 2018-04-10.(b)(6) investigation summary: level a investigation: complaint evaluation / complaint history check for the event(s) that occurred.Severity: s_1__; occurrence: a complaint history check was performed and this is the 1st related complaint for dimension on lot # 8100673 & 8100674.Investigation summary: customer returned (2) open 7mm, 23g pen needles without the tear drop label.Customer states that there is a question on the id measurements.Both returned pen needles were scanned using the ct scanner in order to measure the id on the patient end and non patient end of the cannula of both samples.All measurements fall within specifications and no defects were observed on either of the samples.A review of the device history record was completed for batch# 8100674.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.A review of the device history record was completed for batch# 8100673.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Investigation conclusion: based on the samples and/or photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
|