ARTHREX, INC. FIB REMOVAL KIT (NON-STERILE, DISPOSABLE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
FIB REMOVAL KIT (NON-STERILE, DISPOSABLE |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/07/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
Original dos was (b)(6) 2018, a fibulock nail was used for an ankle fracture.Upon viewing x-rays, the surgeon thought that the nail had broken and so a second surgery was done (b)(6) 2020; he used the cs6910 removal screw from the st6900 fib removal kit.When attempting to remove the nail, the threads of the cs6910 broke off in the nail and then the screw broke.The surgeon determined that the original nail was not broken, so they did not proceed with the removal.The original nail is in the patient; it now contains debris threads from the cs6910.Unknown patient info.
|
| |
|