• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Capturing Problem (2891)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/08/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) had a capture malfunction.The product has been returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: analysis was unable to confirm customer comment of a capture malfunction.Analysis also noted that the hanger assembly was broken, the on/off and enter buttons were out of mechanical specification, the lower case was broken, main seal was pinched, the display was out of electrical specification and the display wires were pinched but wire insulation was not compromised.All defective parts were replaced, and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10011805
MDR Text Key189439189
Report Number3004593495-2020-00492
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169230859
UDI-Public00643169230859
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2020
Initial Date FDA Received04/29/2020
Supplement Dates Manufacturer Received05/12/2020
Supplement Dates FDA Received05/12/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-