It was reported that the ventilator had a dim display and will not shut off or go into standby.There was no patient involvement.The service engineer (se) inspected the device and confirmed the reported issue.The se removed the battery and tried to turn the unit on alternating current (ac) power but the unit would not turn on.The se replaced the power management (pm) board to address the reported issue and the unit was checked overall, run in tested, cleaned, functionally tested and no abnormality was confirmed.
|
B4: 05may2020 g4: (b)(6)2020.The ventilator diagnostic report was provided and reviewed, error codes were found that indicated high tidal volume, low inspiratory pressure, low tidal volume, low minute ventilation, high rate and oxygen not available.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|