| Model Number |
RONYX25038UX |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During a procedure an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a lesion.It was reported that stent deformation occurred in vivo during positioning/advancement.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the device was being used to treat a calcified lesion in the left anterior descending (lad) artery.The device was inspected before use.Negative prep was performed.The lesion was pre-dilated.Resistance was encountered when advancing the device.Excessive force was not used.The device did not pass through a previously deployed stent.The procedure was completed using shorter stents.The patient was reported to be alive with no injury.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information / correction: initial reporters name is unknown.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the stent was positioned correctly at the proximal markerband, however the mid to distal stent struts had stretched over the distal markerband and off the delivery system.Deformation was evident to the mid to distal stent wraps with struts stretched.Deformation was evident to the distal tip.A kink was evident on the distal shaft.The inner lumen patency was verified with a 0.015 inch mandrel.There was no other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|