• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II XR INS XLPE LAT 5-6 LT 8MM; PROSTH, KNEE, PATELLOFEMO, SEMI-CONS, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II XR INS XLPE LAT 5-6 LT 8MM; PROSTH, KNEE, PATELLOFEMO, SEMI-CONS, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 74023860
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Injury (2348)
Date of Event 01/14/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient presented left knee patellar impingement and crepitus.Event was treated with irrigation and debridement.
 
Additional Manufacturer Narrative
It was reported that the patient presented left knee patellar impingement and crepitus.The event was treated with irrigation and debridement.The affected complaint devices, used in treatment, were not returned for evaluation.Therefore a product analysis could not be performed.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident could not be corroborated.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Some potential causes of the reported event could include but not limited to fit / sizing, wear or alignment of device.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, smith and nephew will continue to monitor for future complaints and investigate as necessary.Should additional information be received, this complaint will be reopened and reevaluated.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II XR INS XLPE LAT 5-6 LT 8MM
Common Device Name
PROSTH, KNEE, PATELLOFEMO, SEMI-CONS, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key10014582
Report Number1020279-2020-01417
Device Sequence Number19571414
Product Code JWH
UDI-Device Identifier00885556615744
UDI-Public00885556615744
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152726
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,study,use
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/19/2026
Device Model Number74023860
Device Catalogue Number74023860
Device Lot Number16BM15555
Initial Date Received by Manufacturer 04/06/2020
Supplement Date Received by Manufacturer10/30/2020
Initial Report FDA Received Date04/30/2020
Supplement Report FDA Received Date11/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-