|
SMITH & NEPHEW, INC. MI OFFSET REF IMP SD SHAFT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
71364023 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the surgeon states that this driver when attaching the cup is stripping.No delay to procedure, no injury to patient.S&n backup was available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the hex tip is stripped causing stated failure.The device was manufactured in 2018.The device show significant signs of wear/usage.A review of complaint history on the listed part revealed no prior complaints for the listed batch number with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|
|
|