Model Number 7209485
Device Problem Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/05/2020
Event Type  malfunction  
Event Description
It was reported that, during a meniscus repair surgery, the meniscus mender ii disposable set was found to be damaged before use, a photo shows both loop's heads unsoldered/detached from the shaft.The procedure was successfully completed without significant delay using a back-up device.No patient injuries or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3, h6 one disposable meniscus mender ii set used for treatment, was returned for evaluation.Visual assessment confirmed the complaint.The loop meniscal were broken from the bottom.To avoid components from shifting within the package, individual component storage cradles are intentionally snug.Use of the distal end (head) to remove snare loops from the tray may result in weakening, bending or complete fracture between at the head and shaft connection.Proper method of retrieval is to push and pop the head of the component from the back of the cradle.E1 additional correct information: (b)(6) and address (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET MENISCUS MENDER II DISPOSABLE
Type of Device
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key10015995
MDR Text Key189580650
Report Number1219602-2020-00705
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier03596010468482
UDI-Public03596010468482
Combination Product (y/n)N
PMA/PMN Number
K885311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/22/2024
Device Model Number7209485
Device Catalogue Number7209485
Device Lot Number2031439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2020
Date Manufacturer Received05/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age28 YR
Patient Weight70
-
-