| Model Number |
8890H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a gastric procedure on (b)(6) 2020 and the suture was used.During surgery, the suture broke during tissue passage.There was no harm to patient.The surgery successfully completed.There were no adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent to the fda: 05/27/2020.A manufacturing record evaluation was performed for the finished device batch plj922, and no non-conformances were identified.Additional investigation summary: an opened box with thirteen unopened samples of product code 8890h, lot # plj922 were returned for analysis.The complaint sample was not received for evaluation.During the visual inspection of thirteen unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area of all needles were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the samples received, no breakage suture was found and the tested samples met the finished goods requirements.
|
| |
|
Search Alerts/Recalls
|