• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TISSUE SPARING FEM CORE CASE 1; HIP INSTRUMENTS : INSTRUMENT CASES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TISSUE SPARING FEM CORE CASE 1; HIP INSTRUMENTS : INSTRUMENT CASES Back to Search Results
Model Number 2598-07-390
Medical Device Problem Codes Break (1069); Crack (1135); Material Integrity Problem (2978)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the case has been received for examination.Examination of the returned instrument case confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event or Problem Description
Items are worn and needs to be replaced.Nothing fell into patient.No patient consequence and no surgical delay.Quantities impacted and needing replacement are below (2 and 4 each).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: the case has been received for examination.Examination of the returned instrument case confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TISSUE SPARING FEM CORE CASE 1
Common Device Name
HIP INSTRUMENTS : INSTRUMENT CASES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10016083
Report Number1818910-2020-11575
Device Sequence Number16969932
Product Code FSM
UDI-Device Identifier10603295148333
UDI-Public10603295148333
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2598-07-390
Device Catalogue Number259807390
Device Lot NumberC15AP4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2018
Initial Date Received by Manufacturer 04/23/2020
Supplement Date Received by Manufacturer04/23/2020
Initial Report FDA Received Date04/30/2020
Supplement Report FDA Received Date05/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-