This report summarizes 9 malfunction events.A review of the events indicated that eight (8) patient sample tests using the capture-r ready id reagent on automated blood bank systems produced nine (9) unexpected negative results for specific antibodies.The results represent false negatives due to prior patient histories or, testing completed by other methods that demonstrated the presence of the specific antibodies.The malfunctions did not show positive results as expected for the anti-e antibody on one (1) occasion; the anti-e antibody on three (3) occasions; the anti-k antibody on one (1) occasion; the anti-m antibody on one (1) occasion; the anti-fya on two (2) occasions and anti-c on one (1) occasion.Subsequent testing of reagent retention lots, reviews of design history records and reagent antigen validation testing of the initial bulk product used for commercial vialing showed acceptable results.Through post event tests and investigations, no specific causes were determined for the malfunctions.
|