• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE R READY SCREEN POOLED; REAGENT RED BLOOD CELLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE R READY SCREEN POOLED; REAGENT RED BLOOD CELLS Back to Search Results
Catalog Number 006436
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
This report summarizes 3 malfunction events.A review of the events indicated that three (3) patient sample tests using the capture-r ready screen pooled reagent on automated blood bank systems produced unexpected negative results for specific antibodies.The results represent false negatives due to prior patient histories or, testing completed by other methods that demonstrated the presence of the specific antibodies.The malfunctions did not show positive results as expected for the anti-m antibody on one (1) occasion; the anti-d antibody on one (1) occasion and the anti-jka antibody on one (1) occasion.Subsequent testing of reagent retention lots, reviews of design history records and reagent antigen validation testing of the initial bulk product used for commercial vialing showed acceptable results.Through post event tests and investigations, no specific causes were determined for the malfunctions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE R READY SCREEN POOLED
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross, ga
MDR Report Key10016378
Report Number1034569-2020-00034
Device Sequence Number2462179
Product Code QHT
UDI-Device Identifier10888234001959
UDI-Public10888234001959
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
102707/0.0
Number of Events Summarized3
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number006436
Device Lot NumberN584.N588,N59204582
Was Device Available for Evaluation? Yes
Type of Report(Section G)Periodic
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-