| Model Number |
MMT-523RNAH |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump passed the rewind, basic occlusion, prime/a33 and displacement test.Pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had motor error alarm.Customer was able to successfully clear the alarm and able to rewind the insulin pump.Customer reported that drive support cap was normal and insulin pump had not been exposed to high magnetic field.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|