ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number ASKU
Device Problem Device Alarm System (1012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/17/2020
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported there was no system message displayed when the saline solution bottle was emptied during the viscoelastic removal portion of a cataract extraction with intraocular lens implant procedure.The staff reported a bad connection of the hoses on the cassette which could have become loosened due to not being tightened properly.There was air in the irrigation line and the device aspirated part of the zonular fibers and vitreous appeared.A capsular tension ring was placed and an anterior vitrectomy was performed.
 
Manufacturer Narrative
Additional information has been provided in h.6.And h.10.No further information was able to be obtained from this customer.With no additional, related information provided, the customers reported event was not able to be confirmed.Posterior capsule tear is an issue that is occasionally reported with cataract surgery.A root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10020118
MDR Text Key189695588
Report Number2028159-2020-00369
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/17/2020
Initial Date FDA Received05/04/2020
Supplement Dates Manufacturer Received07/24/2020
Supplement Dates FDA Received08/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-