The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on (b)(6) 2020, the two (2) trays; matrix mandible and matrix midface tray were inspected because the trays had fallen from a shelf in the sterile processing department (spd).Several parts were noted to be damaged in the tray.One (1) bone reduction forcep had a broken fine tip, one (1) screwdriver handle was broken end cap is missing, one (1) matrix mandible drill guide broken tip damage on which the drill bit would not easily sliding through, and one (1) t-handle was broken slide is very loose.There was no patient involvement.This is report 2 of 4 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part number:311.006, synthes lot number: 5341253, supplier lot number: lm016190, release to warehouse date: september 20, 2006, expiration date: n/a, supplier: (b)(4).No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Visual inspection: the screwdriver handle with hex coupling-medium (p/n: 311.006, lot number: lm016190) was received at us customer quality (cq).Upon visual inspection, the shaft clamp was missing.Additionally, the flanged bearing was damaged with small pieces missing.Device failure/defect identified? yes.Dimensional inspection: no dimensional inspection was performed due to post-manufacturing damage and definitive finding of a missing component.Document/specification review: current and manufactured were reviewed.No design issues or discrepancies were identified.Complaint confirmed? yes.Investigation conclusion: this complaint is confirmed as the shaft clamp was missing.Additionally, the flanged bearing was damaged with small pieces missing.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|