| Model Number |
52-1485 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377); No Code Available (3191)
|
| Date of Event |
05/04/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Der states asr revision, originally implanted by the surgeon.Cobalt level of 4.4, chromium level of 4.7.Srom stem stayed implanted.Revised cup to a competitive cup, replaced srom head.No further patient information available.Doi: (b)(6) 2008; dor: (b)(6) 2018; right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : the complaint investigation concluded that the event was related to the asr platform which was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.This stem is not part of the asr family however it did form part of the joint construct.There was no allegation of deficiency with this device.
|
| |
|