Catalog Number UNKNOWN |
Device Problems
Structural Problem (2506); Detachment of Device or Device Component (2907)
|
Patient Problems
Device Embedded In Tissue or Plaque (3165); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturer ref# (b)(4).Catalog# is unknown but referred to as cook gunther tulip filter.Occupation: non-healthcare professional.Pma/510(k) k172557.Summary of investigational findings: the reported allegations have been investigated based on the information provided to date.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Cook medical will continue to monitor for similar events.
|
|
Event Description
|
Description of event according to initial reporter: it is alleged that "on or about (b)(6) 2010 [pt] was implanted with a cook gunther tulip filter prior to lumbar spinal procedure.In (b)(6) 2020, it was discovered that the cook filter had fractured and [pt] is considering his medical options to treat the cook filter failure.Patient outcome: [pt] is at risk for future cook filter fractures, migrations of the filter or embolized fragments, embedment, and/or tilt.He faces numerous health risks, including the risk of death.For the rest of [pt]'s life, he will require on-going medical monitoring and use of anti-coagulants.".
|
|
Manufacturer Narrative
|
Investigation: the following allegations have been investigated: embedded.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: vascular trauma, vena cava perforation, vena cava penetration.Catalog number and lot number are unknown; however, the alleged filter is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Filter was allegedly implanted via the left common femoral vein.(b)(6) 2010, per a report from implant report; ¿the filter was inserted and deployed at the l2 level.¿ (b)(6) 2020, per a report from angiogram; ¿suspected portion of fractured inferior vena cava filter located against the interatrial septum on the right ventricular side.There is also what appears to be a fractured hook/limb of the ivc filter seen in the vena cava.¿.
|
|