• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Difficult to Remove (1528); Structural Problem (2506); Insufficient Information (3190)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Internal Organ Perforation (1987); Perforation of Vessels (2135); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Manufacturer ref# (b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2010.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
 
Event Description
Patient allegedly received an implant via the right internal jugular due to deep vein thrombosis and pulmonary embolism.Patient is alleging vena cava perforation, embedment, complicated removal.Patient further alleges pain, physical limitations, depression.Report from computerized tomography (ct): "ivc filter is present.Some of the distal struts of the filter extend outside the lumen of the inferior vena cava, and this appears increased since prior 2011ct.This is greatest involving one of the anterior struts which extends 1.6cm outside of the anterior wall of the inferior vena cava." report from ct: "an ivc filter is demonstrated with prongs extending outside of the ivc lumen, most prominently in the left lateral projections where there is approximately 1.5 cm and 1 cm of extrusion respectively.The anterior prong likely extends into the duodenum without complications." retrieval report (successful): "attempts made to grasp the apex of the filter with the snare, however these were unfruitful as the filter apex was incorporated into the wall." " the ivc filter was subsequently removed.".
 
Manufacturer Narrative
Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up.The risks/benefits of filter retrieval should be considered for each patient during follow-up.Once protection from pe is no longer necessary, filter retrieval should be considered.Filter retrieval should be attempted when feasible and clinically indicated.Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e.G., a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications).For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set.It may also be retrieved with the cloversnare® vascular retriever.Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques.The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position.The safety or effectiveness of these alternative retrieval techniques has not been established.Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve.For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.Unknown if the reported pain, physical limitations, depression are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Rpn and lot# are unknown the following allegations have been investigated: vena cava/organ perforation, embedment, complicated removal, pain, physical limitations, depression.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key10023160
MDR Text Key189823902
Report Number3002808486-2020-00519
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/27/2020
Initial Date FDA Received05/05/2020
Supplement Dates Manufacturer Received05/20/2020
Supplement Dates FDA Received06/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight122
-
-