• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED XMTR MMT-7811NA GST3C LOCKOUT; SENSOR, GLUCOSE, INVASIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED XMTR MMT-7811NA GST3C LOCKOUT; SENSOR, GLUCOSE, INVASIVE Back to Search Results
Model Number MMT-7811NA
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was able to charge, communicated and sync properly during radio frequency association.However, device failed accuracy test with isig values out of range; problem was isolated to electronic assembly.
 
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic the user reported inaccurate sensor readings that triggered a threshold suspend.The customer¿s blood glucose was 112 mg/dl and sensor glucose was 41 mg/dl at the time of the event, the difference is outside the acceptable range.Suspend on low limit in sensor settings was at 41 mg/dl.No harm requiring medical intervention was reported.The sensor will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XMTR MMT-7811NA GST3C LOCKOUT
Common Device Name
SENSOR, GLUCOSE, INVASIVE
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key10023852
Report Number2032227-2020-113221
Device Sequence Number11502163
Product Code MDS
UDI-Device Identifier00763000215538
UDI-Public(01)00763000215538(17)200630
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2020
Device Model NumberMMT-7811NA
Device Catalogue NumberMMT-7811NA
Device Lot NumberA000769495
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2020
Initial Date Received by Manufacturer 04/30/2020
Supplement Date Received by Manufacturer05/13/2020
Initial Report FDA Received Date05/05/2020
Supplement Report FDA Received Date05/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
Patient Weight170
-
-