| Model Number |
10013890 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that during an infusion the tubing appeared to have micro holes where the fluid appeared to seep from the tubing where the cartridge is inserted into the pump.It was reported no medical treatment was required as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report that during an infusion the tubing appeared to have micro holes where the fluid appeared to seep from the tubing where the cartridge is inserted into the pump was confirmed.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.A small hole was observed in the silicone tubing just below the upper fitment retainer ring.No gouge/crush marks were observed on the upper fitment.Clear liquid was observed in the set¿s drip chamber and entire length of tubing.No other abnormalities were observed with the set.Functional testing was performed.Blue-dyed water was observed to be dripping out of the location of the hole in the silicone segment tubing.The seeping of fluid was due to a hole in the silicone tubing just below the upper fitment retainer ring.The root cause of the hole could not be definitively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that during an infusion the tubing appeared to have micro holes where the fluid appeared to seep from the tubing where the cartridge is inserted into the pump.It was reported no medical treatment was required as a result of the event.
|
| |
|