| Model Number |
MMT-551LNAH |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Power Problem (3010); Patient Device Interaction Problem (4001); Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device monitored for several days.Unit received with all operating currents within specification and passed functional testing including the rewind, a21 error test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.Device primed properly.No excessive no delivery/occlusion alarm noted.No unexpected failed battery alarm anomalies noted.Device received with cracked battery tube threads and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump sometimes shoots or squirts insulin from infusion set during priming.Customer also stated that insulin pump had rejected new batteries, had cracks in insulin pump casing near battery threading and had unexplained or random no delivery alarms no harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|