• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMÉRIEUX SA CHROMID® S.AUREUS ELITE; CHROMID S.AUREUS ELITE 20 PL - 419042

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMÉRIEUX SA CHROMID® S.AUREUS ELITE; CHROMID S.AUREUS ELITE 20 PL - 419042 Back to Search Results
Catalog Number 419042
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
On 16-apr-2020, a (b)(6) customer reported absence of typical pink coloration when growing s.Aureus strains with chromid s.Aureus elite 20 pl, reference 419042, lot 1007883410, expiry date 29-may-2020.The customer reported that typical pink colored colonies for a s.Aureus patient strain were not observed as expected.The same strain showed pink colonies on a lot number previously used by the customer.The customer subsequently subcultured the same s.Aureus patient strain on two different lot numbers of reference 419042 and on blood agar in parallel.On lot 1007883410, colonies are beige-yellow and on a different lot number, colonies are pink-purple as expected for s.Aureus.The customer also observed this issue on others samples (blood cultures or orthopedic wound samples).There is no indication from the customer that this event impacted the patient state of health or led to a delay of results.A biomérieux internal investigation has been initiated.
 
Additional Manufacturer Narrative
This report was initially submitted following notification from a customer in france regarding the absence of typical pink coloration when growing s.Aureus strains with chromid s.Aureus elite 20 pl (reference 419042, lot 1007883410).An internal investigation was conducted using biomerieux retained samples of the subject lot, and on the plates received from the customer.The testing results confirmed the product performed within expected specifications.The customer issue was not reproduced through the internal investigation performed on the retain samples, nor with the customers plates provided.A root cause of the event could not be established.See section h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHROMID® S.AUREUS ELITE
Common Device Name
CHROMID S.AUREUS ELITE 20 PL - 419042
Manufacturer (Section D)
BIOMÉRIEUX SA
5, rue des aqueducs
craponne 69290
FR  69290
MDR Report Key10028243
Report Number9615755-2020-00006
Device Sequence Number19002380
Product Code JSI
UDI-Device Identifier03573026494445
UDI-Public03573026494445
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
C1 EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/29/2020
Device Catalogue Number419042
Device Lot Number1007883410
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2020
Supplement Date Received by Manufacturer05/21/2020
Initial Report FDA Received Date05/06/2020
Supplement Report FDA Received Date06/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-