| Model Number |
MMT-723RNAS |
| Medical Device Problem Codes |
No Display/Image (1183); Device Difficult to Program or Calibrate (1496); Moisture or Humidity Problem (2986); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the customer went for swimming and now insulin pump won't work.The customer stated that the insulin pump had button error.The customer also stated that the minutes are advancing.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to moisture damage to the electronic devices note.
|
| |
|