| Model Number |
370946 |
| Medical Device Problem Codes |
Failure to Capture (1081); Ambient Noise Problem (2877); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Dizziness (2194)
|
| Date of Event |
04/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a decrease in impedance detected on this rv lead.The patient experienced slight dizziness.Lead check revealed noise and failure to capture.The lead was explanted and replaced.
|
| |
|