MAUDE Adverse Event Report: ACE SURGICAL SUPPLY COMPANY TRICAM INTERNAL IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACE SURGICAL SUPPLY COMPANY TRICAM INTERNAL IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number 305011
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Codes Failure of Implant (1924); Swelling (2091)
Date of Event 03/18/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load failure.This implant loss suggests that the source of the problem was likely to stem from no osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, overall health or behavior.Proper intended use of the implant is described in its instructions for use.
 
Event or Problem Description
Mobility and swelling were reported.Bone quality at the time of implant failure type iii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability and osseointegration was not achieved.Augmentation procedure was not performed at the time of surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICAM INTERNAL IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton, ma
MDR Report Key10032720
Report Number1287163-2020-00726
Device Sequence Number3116591
Product Code DZE
UDI-Device Identifier00614950005284
UDI-Public00614950005284
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K103790
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number305011
Device Catalogue Number305011
Device Lot Number19040051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/10/2020
Initial Report FDA Received Date05/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-