MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problems Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Patient Problems Hyperglycemia (1905); Unspecified Infection (1930)
Event Date 03/27/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the customer was hospitalized due to sever infection at the infusion site and also had experienced high blood glucose.The customer¿s blood glucose level was over 470 mg/dl.It was reported that the customer hospitalized from (b)(6) 2020.It was reported that the customer was treated in hospital for infection.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device passed the functional tests, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and delivery accuracy test at 0.08710 inches.Device received with scratched case, pillowing keypad overlay and minor scratched lcd window.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key10032755
MDR Text Key190139965
Report Number2032227-2020-114762
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00763000190439
UDI-Public(01)00763000190439(17)211203
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/03/2021
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG30DSM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received05/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/04/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN- MMT332-RSVR, UNOMED SET
Patient Outcome(s) Hospitalization;
Patient Age65 YR
-
-