MDT SOFAMOR DANEK PUERTO RICO MFG CAPSTONE CONTROL PTC SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number 5011122
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient presented with lumbar canal stenosis; and underwent transforaminal lumbar interbody fusion of 3 intravertebral discs (l3-s1).Intra-op, after inserting the cage at l4-l5, there was a sense of incongruity when rotating the cage.When the cage was removed, it was found that the cage had been twisted and deformed.Also, the inserter's tip was found to be widened, the surgeon placed it back with pliers and the inserter was continued to be used.The deformed cage was replaced with a new one to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Device evaluation summary: visual and optical inspection confirmed the implant has been damaged.The threads have been stripped and the implant appears to be bent and twisted.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSTONE CONTROL PTC SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key10032778
Report Number1030489-2020-00506
Device Sequence Number16769764
Product Code MAX
UDI-Device Identifier00643169677517
UDI-Public00643169677517
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K171107
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5011122
Device Catalogue Number5011122
Device Lot NumberH5533802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2020
Supplement Date Received by Manufacturer08/21/2020
Initial Report FDA Received Date05/07/2020
Supplement Report FDA Received Date08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INSERTER; INSERTER
Patient Age77 YR
Patient Weight55
-
-