Model Number C304-HIS
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during an implant procedure the catheter was difficult to slit and the core wire was exposed.The catheter was used to complete the procedure.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the delivery catheter was returned without the dilator and analyzed.Analysis indicated the mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.There was an issue with the catheter wire.The mechanical operation of the catheter shaft was kinked/buckled.Visual analysis of the lead indicated damage during use.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELECTSITE
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10033486
Report Number9612164-2020-01810
Device Sequence Number3116671
Product Code DQY
UDI-Device Identifier00763000147006
UDI-Public00763000147006
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K183493
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model NumberC304-HIS
Device Catalogue NumberC304-HIS
Device Lot Number0009995971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2020
Initial Date Received by Manufacturer 04/16/2020
Supplement Date Received by Manufacturer05/11/2020
Initial Report FDA Received Date05/07/2020
Supplement Report FDA Received Date05/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
-
-