(b)(4).Investigation results: visual examination of the returned complaint device under high magnification found the balloon had a pinhole in the distal section over the ro marker.Dry contrast was found inside the balloon.Functional analysis was attempted to be performed; however, the balloon lumen was occluded and could not be unblocked.This failure is likely due to factors encountered during the procedure, such as handling of the device, the technique used by the physician, the interaction with the scope, and normal procedural difficulties encountered during the procedure, that could have affected device performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
|
Note: this report pertains to one of two hurricane rx dilation balloons used during the same procedure.It was reported to boston scientific corporation that the two hurricane rx dilation balloons were used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, it was noted that the two balloons did not inflate at all.The procedure was completed with a different device.There were no patient complications reported as a result of this event.Investigation results revealed that the balloon had a pinhole; therefore, this is now an mdr reportable event.
|