• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION HURRICANE RX; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION HURRICANE RX; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00545890
Medical Device Problem Codes Material Puncture/Hole (1504); Obstruction of Flow (2423)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation results: visual examination of the returned complaint device under high magnification found the balloon had a pinhole in the distal section over the ro marker.Dry contrast was found inside the balloon.Functional analysis was attempted to be performed; however, the balloon lumen was occluded and could not be unblocked.This failure is likely due to factors encountered during the procedure, such as handling of the device, the technique used by the physician, the interaction with the scope, and normal procedural difficulties encountered during the procedure, that could have affected device performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event or Problem Description
Note: this report pertains to one of two hurricane rx dilation balloons used during the same procedure.It was reported to boston scientific corporation that the two hurricane rx dilation balloons were used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, it was noted that the two balloons did not inflate at all.The procedure was completed with a different device.There were no patient complications reported as a result of this event.Investigation results revealed that the balloon had a pinhole; therefore, this is now an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HURRICANE RX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10035141
Report Number3005099803-2020-01849
Device Sequence Number2986107
Product Code FGE
UDI-Device Identifier08714729283805
UDI-Public08714729283805
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K001338
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberM00545890
Device Catalogue Number4589
Device Lot Number0022727607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/13/2020
Initial Report FDA Received Date05/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-