| Model Number |
401729 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The orsiro drug-eluting stent system was chosen for the treatment of the target lesion in the lad.During lesion crossing the stent dislodged from the balloon but was successfully captured and dilated over the wire using different balloon.Eventually the stent deployed at undesired location.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|
Search Alerts/Recalls
|