Model Number APB-1-2-HX-ES
Medical Device Problem Codes Separation Failure (2547); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the axium prime coil would not detach, and then prematurely detached upon removal.It was reported that the spring coil couldn¿t be detached when implanted.The physician attempted to detach the coil twice manually.During the process of pushing it out of the body, it automatically detached.The coil was removed from the patient and replaced with another medtronic coil.The pushwire was not bent or broken.There was no friction or difficulty during delivery.The physician did not reposition the coil or rotate the delivery pusher during the procedure.The patient did not experience any injury or complications.The devices were prepared according to the instructions for use (ifu).A continuous flush was administered during the procedure.The patient was undergoing treatment for a unruptured, saccular aneurysm located in the left middle cerebral artery branch between m1 and m2.The max diameter was 4.58 mm.The neck diameter was 3.97 mm.The patient¿s blood flow and tortuosity were normal.The patient¿s medical history included smoking and hypertension for more than 10 years.Ancillary devices include a navien guide catheter and an echelon microcatheter.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BRPL HLX
Common Device Name
DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10037161
Report Number2029214-2020-00437
Device Sequence Number16769809
Product Code KRD
UDI-Device Identifier00847536032095
UDI-Public00847536032095
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151447
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/17/2022
Device Model NumberAPB-1-2-HX-ES
Device Catalogue NumberAPB-1-2-HX-ES
Device Lot NumberA933866
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2020
Supplement Date Received by Manufacturer05/11/2020
Initial Report FDA Received Date05/07/2020
Supplement Report FDA Received Date05/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
Patient Weight65
-
-