| Catalog Number |
315-07-06 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
As reported, during a procedure on a (b)(6) y/o male patient, the surgeon impacted the glenoid component with the plastic black tip, it fractured in half.The broken impactor tip pieces were removed from the wound.The patient was last known to be in stable condition following the event.The device is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot number: 60383001.(h3) based on historical complaint investigations and the returned device, the fractured device reported was likely the result of crack initiation, propagation, and ultimate fracture of the device which was likely caused by abnormal or aggressive use inconsistent with the intended use of this device.
|
| |
|