MEDTRONIC ADVANCED ENERGY (SALIENT) PLASMABLADE NEEDLE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number PS200-001
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a manufacture representative (rep) regarding a handpiece.It was reported that the handpiece tip broke off when cleaning the distal tip of the needle with a wet gauze.This broke at the end of the procedure and as a result of this, nothing has happened to the patient.
 
Additional Manufacturer Narrative
Analysis summary: complaint confirmed: upon receiving the device, the sender provided the broken needle from the device in a cup for the analysis.The heat shrink on the shaft was cut to expose and investigate the weld from the needle tip.After observing the welding of the shaft to the needle tip, it was determined that the contributing factor was due to a weak weld during the welding phase in assembly.No functional tests were required.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLASMABLADE NEEDLE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MEDTRONIC ADVANCED ENERGY (SALIENT)
180 international drive
portsmouth NH 03801
MDR Report Key10038733
Report Number1226420-2020-00061
Device Sequence Number2501818
Product Code GEI
UDI-Device Identifier00643169628410
UDI-Public00643169628410
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K082786
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/12/2021
Device Model NumberPS200-001
Device Catalogue NumberPS200-001
Device Lot Number0216566093
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2020
Supplement Date Received by Manufacturer06/05/2020
Initial Report FDA Received Date05/08/2020
Supplement Report FDA Received Date06/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No