Model Number 96-6115 |
Device Problem
Device-Device Incompatibility (2919)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Match issue.It was reported that as the event (b)(4) happened, considering safety, the distributor checked the same lot products, noted the impact product also with the same issue: 966112 could not pass the way through the slot/hole of 966115.There were no adverse consequences to the patient.No additional information could be provided.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirms the reported event.Root cause and corrective action are documented in the capa system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Manufacturer Narrative
|
Product complaint # = > (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added d11 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Search Alerts/Recalls
|