MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX PEEK ANCH.W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX PEEK ANCH.W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 222209
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Initial reporter phone number (b)(6).Udi: (b)(4).
 
Event or Problem Description
It was reported that during the surgery of medial collateral ligament repair, when inserting the anchor 4.5 healix peek anch.W/ocord, the anchor was broken off.This issue re-occurred on 222209(5.5 healix peek anch.W/ocord).Another device was used to complete the surgery.There were no adverse consequences to the patient.The devices are available to be returned for evaluation.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: b5; subsequent follow-up with the customer, additional information was received.It was reported that there was no surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> according to the information provided, it was reported that during the surgery of medial collateral ligament repair, when inserting the anchor 4.5 healix peek anch.W/ocord, the anchor was broken off.This issue re-occurred on 222209(5.5 healix peek anch.W/ocord).The complaint device was received and evaluated.Visual observation confirms that the distal tip of the anchor is broken and is held in place by suture.When observed under magnification, no structural anomalies were observed that could have contributed to this failure.The complaint can be confirmed.The possible root cause for reported failure can be associated with off axis insertion, levering during insertion, hard bone quality or using incorrect instrumentation for preparing bone hole.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX PEEK ANCH.W/OCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10040599
Report Number1221934-2020-01341
Device Sequence Number14187298
Product Code MAI
UDI-Device Identifier10886705007578
UDI-Public10886705007578
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number222209
Device Catalogue Number222209
Device Lot Number4L88548
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2020
Initial Date Received by Manufacturer 04/26/2020
Supplement Date Received by Manufacturer05/10/2020
05/20/2020
05/29/2020
Initial Report FDA Received Date05/08/2020
Supplement Report FDA Received Date05/11/2020
05/26/2020
05/29/2020
Patient Sequence Number1
Patient Age48 YR
-
-