• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KR 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KR 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KR
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fatigue (1849); Hyperglycemia (1905)
Event Date 05/06/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call that they experienced high blood glucose of 436 mg/dl.Customer¿s other blood glucose was 400 mg/dl.The customer did experienced the symptoms such as fatigue.The customer treated with the manual injection.Troubleshooting was done for high blood glucose and under delivery.The drive support cap appears normal.The customer was neither in the emergency room, nor admitted into hospital as a result of high blood glucose.Customer reported that cannula was bent.Based on customer report customer did not allege insulin pump was under delivering and customer was performing high pressure test and it was passed.The insulin pump will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KR 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
8185464805
MDR Report Key10042248
MDR Text Key190474100
Report Number2032227-2020-116197
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00763000173036
UDI-Public(01)00763000173036
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-1715KR
Device Catalogue NumberMMT-1715KR
Device Lot NumberHG35527
Was Device Available for Evaluation? No
Date Manufacturer Received05/06/2020
Date Device Manufactured01/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-MMT-326-RSVR,UNOMED SET
Patient Outcome(s) Other;
Patient Age77 YR
-
-