MICRO THERAPEUTICS, INC. DBA EV3 ONYX AVM; AGENT, INJECTABLE, EMBOLIC Back to Search Results
Model Number 105-7100-080
Medical Device Problem Code Component or Accessory Incompatibility (2897)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding onyx solution that could not be visualized well during a procedure.The patient was being for a grade 4 arteriovenous malformation.Vessel tortuosity was severe.It was reported that the onyx solution was prepared according to the manufacturer instructions for use.It was shaken for over 60 minutes and was used for injection immediately after mixing.Upon injecting the onyx solution, it was not visualized exiting the catheter nor could it be seen in the catheter.The catheter was inspected for possible rupture but none was observed.There was no harm or injury to the patient.
 
Additional Manufacturer Narrative
H6.Codes updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reporting that the avm was located on the patient's rightside involving the external carotid and middle cerebral arteries.The onyx injection was continuous; the physician did not pause.The dead space of the catheter was filled with dimethyl sulfoxide (dmso).There was no further surgical or medical intervention required due to the issue.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX AVM
Common Device Name
AGENT, INJECTABLE, EMBOLIC
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10042572
Report Number2029214-2020-00443
Device Sequence Number2366653
Product Code MFE
UDI-Device Identifier00847536006188
UDI-Public00847536006188
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2022
Device Model Number105-7100-080
Device Catalogue Number105-7100-080
Device Lot NumberA822862
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2020
Supplement Date Received by Manufacturer05/07/2020
05/07/2020
Initial Report FDA Received Date05/11/2020
Supplement Report FDA Received Date06/22/2020
06/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age47 YR
-
-