Model Number TABLETOP-JAPAN |
Device Problems
Failure to Shut Off (2939); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A surgeon reported that during a surgical procedure the system froze.The staff tried to reboot however the device was unable to turn off.The power cord was unplugged to turn the power off.The device was then rebooted and the surgery was completed.There was a one and a half hour delay but no patient harm was reported.Additional information has been requested.
|
|
Manufacturer Narrative
|
The company service representative examined the system and was able to confirm and replicate the reported event.The nonconforming host module was replaced.The power module was replaced as a preventative measure (pm).The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to a nonconforming host module.The manufacturer internal reference number is: (b)(4).
|
|