ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP-JAPAN
Device Problems Failure to Shut Off (2939); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/24/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A surgeon reported that during a surgical procedure the system froze.The staff tried to reboot however the device was unable to turn off.The power cord was unplugged to turn the power off.The device was then rebooted and the surgery was completed.There was a one and a half hour delay but no patient harm was reported.Additional information has been requested.
 
Manufacturer Narrative
The company service representative examined the system and was able to confirm and replicate the reported event.The nonconforming host module was replaced.The power module was replaced as a preventative measure (pm).The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to a nonconforming host module.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10043095
MDR Text Key190774719
Report Number2028159-2020-00376
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-JAPAN
Device Catalogue Number8065751817
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-