Drive devilbiss is the initial importer of the device which is a knee walker.This report is being tendered in an overabundance of caution in response to an mdr regression analysis.The device's seat post failed which caused the customer to fall.She hit her head and received a minor concussion.Since this incident a design change under pc#297 (2016/2017) has addressed the seat post issue.The serial number provided shows that the unit was manufactured in 2013, prior to the changes.
|