U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 9 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FB TIB BASE SZ 9 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Model Number 1506-00-009
Medical Device Problem Codes Loss of or Failure to Bond (1068); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 11/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient received right primary attune tka to treat osteoarthritis with a mild varus deformity and significant medial compartment wear.The patella was resurfaced and depuy cement x 2 was utilized.The procedure was completed without complications.Patient received right knee revision to treat pain secondary to loosening of the tibial tray.Upon entering the joint, the surgeon confirmed tibial tray loosening at an unknown interface.The femoral component was well-fix but revised.The surgeon notes there was a large cyst in the posterior distal portion of the femoral condyle.The patella was well-fixed and retained.There was no reported product problem with the explanted tibial insert.The patient was revised with competitor products.The procedure was completed without complications.Doi: (b)(6) 2014.Dor: (b)(6) 2018; right knee.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 9 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10044714
Report Number1818910-2020-11956
Device Sequence Number19814315
Product Code JWH
UDI-Device Identifier10603295042433
UDI-Public10603295042433
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number1506-00-009
Device Catalogue Number150600009
Device Lot Number7829930
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/28/2020
Supplement Date Received by Manufacturer06/09/2020
Initial Report FDA Received Date05/11/2020
Supplement Report FDA Received Date06/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 9 CEM; ATTUNE MEDIAL DOME PAT 38MM; ATTUNE PS FB INSRT SZ 9 10MM; ATTUNE PS FEM RT SZ 9 CEM; DEPUY CMW 1 40G; DEPUY1 GENTAMICIN CEMENT 40G; ATTUNE FB TIB BASE SZ 9 CEM; ATTUNE MEDIAL DOME PAT 38MM; ATTUNE PS FB INSRT SZ 9 10MM; ATTUNE PS FEM RT SZ 9 CEM; DEPUY CMW 1 40G; DEPUY1 GENTAMICIN CEMENT 40G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-