| Model Number |
ME1110 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.No additional information was available.
|
| |
|
Event or Problem Description
|
|
Unexpected return - it was reported from the nicu that the iv tubing was missing the male luer.Per customer, no additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported of iv tubing missing male luer was confirmed and identified to be a cracked male luer hub.It was observed that the received set sample's male luer collar was damaged.No testing was deemed necessary due to the obvious damage.Device history record for model me1110 lot 19106625 shows that the set was manufactured on 19 october 2019 with a total of 12,003 units.There were no qn¿s (quality notification) issued during the production build of this lot for the failure mode reported.The root cause was design of the male luer component.
|
| |
|
Event or Problem Description
|
|
Unexpected return - it was reported from the nicu that the iv tubing was missing the male luer.Per customer, no additional information is available.
|
| |
|