It was reported that during an angioplasty procedure in the brachial artery, the pta balloon allegedly ruptured.It was further reported that the balloon tore in half after the delivery system became stuck on the guidewire.Reportedly, medical intervention was required to remove the detached segment.The detached segment was successfully removed from the patient.Patient is reportedly stable.
|
|
Manufacturing review: a lot history review was conducted and it was determined that a device history record (dhr) review was not required.Investigation summary: the device was returned for evaluation.A visual inspection was performed, and the device was found in two segments.Segment one consist of the partial balloon (circumferential rupture) and catheter shaft.Segment two consist of the partial outer catheter shaft and remaining balloon.No other anomalies were noted to the device.Performance evaluation was not performed due to the condition of the device.Therefore, the investigation is confirmed for the reported detachment of device or device component, as the device was returned in two segments, with a portion of the balloon and catheter shaft and also confirmed for the reported balloon rupture, as the circumferential balloon rupture was noted during visual inspection.The definitive root cause for the reported material rupture and detachment of device or device component could not be determined based upon available information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 05/2022),g4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that during an angioplasty procedure in the brachial artery, the pta balloon allegedly ruptured.It was further reported that the balloon tore in half after the delivery system became stuck on the guidewire.Reportedly, medical intervention was required to remove the detached segment.The detached segment was successfully removed from the patient.Patient is reportedly stable.
|