• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL 5.5 HEALIX ADVANCE KNTLS PEEK; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL 5.5 HEALIX ADVANCE KNTLS PEEK; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 222585
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
It was reported by the sales rep via phone that during a rotator cuff repair procedure the tip of this 5.5 healix advance knotless peek broke.No fragments generated.Another device was used to complete the procedure.No patient consequences or surgical delay reported.The device was discarded by the staff.Additional information provided by the affiliate reported the device broke when pulling sutures through the metal kite, the tip of the anchor was broken.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Th excessive force applied while pulling sutures through the metal kite would have caused the tip of the anchor to break, therefore is a possible root cause of the reported problem.However without the return of the complaint device, and no further information provided we cannot determine a definitive root cause.A manufacturing record evaluation was performed for the finished device lot number (l918902), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot : a manufacturing record evaluation was performed for the finished device lot number (l918902), and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX ADVANCE KNTLS PEEK
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10049259
Report Number1221934-2020-01358
Device Sequence Number3292720
Product Code MBI
UDI-Device Identifier10886705030347
UDI-Public10886705030347
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K171114
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Model Number222585
Device Catalogue Number222585
Device Lot NumberL918902
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/27/2020
Supplement Date Received by Manufacturer05/27/2020
Initial Report FDA Received Date05/12/2020
Supplement Report FDA Received Date05/28/2020
Patient Sequence Number1
-
-