Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and delivery accuracy test at 0.08740 inches.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device uploaded properly using carelink.Device had cracked case at display window corner, cracked battery tube threads and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.(b)(4).
|